LABORATORY EQUIPMENT
Bought - Sold - Exchanged




< Prev   INCUBATORS   Next >

Item 3172 INC - Forma 3033 CO2 SteriCult 200

3172 INC

Thermo Stericult 3033

The Thermo Forma Steri-Cult CO2 3033 Incubator uses HEPA filtered, vertical laminar airflow techniques and a filtered external water supply to minimize contamination and desiccation of cultures. The Thermo Forma 3033 has an innovative humidity supply system located outside the chamber that provides rapid RH recovery. The humidity system is not a vapor generator and has no immersible heater to burn out or to cause scaling. Water reservoir for the humidity system is easily filled by way of a port on the outside of the unit. Humidity control is user-selectable above ambient for total control over chamber conditions.

All of our equipment is used and reconditioned. We test CO2 incubators using a calibrated digital chart recorder to verify the accuracy of temperature controls and readouts, as well as using a Fyrite indicator to measure CO2 levels in the chamber. We replace any hardware or parts as is necessary and clean all incubators thoroughly prior to shipment.

SPECIFICATIONS

Temperature

Control . . . . . . . . . . . . . . . ±0.1C

Range . . . . . . . . . . . . . . . . 4C above ambient to 45C (113F)

Sensor . . . . . . . . . . . . . . . . Platinum

Controller . . . . . . . . . . . . . Microprocessor

Setpoint . . . . . . . . . . . . . . . Digital

Display . . . . . . . . . . . . . . . Digital LCD

Readability . . . . . . . . . . . . 0.1C

Setability . . . . . . . . . . . . . . 0.1C

Uniformity. . . . . . . . . . . . . ±0.3C @ 37C (98.6F)

Temperature Alarm

Sensor . . . . . . . . . . . . . . . . Platinum

Controller . . . . . . . . . . . . . Independent analog (high and low alarm)

Setpoint . . . . . . . . . . . . . . . Digital with screwdriver adjust

Display . . . . . . . . . . . . . . . Audible/visual

Readability . . . . . . . . . . . . 0.1C

Setability . . . . . . . . . . . . . . 0.1C

CO2

Control . . . . . . . . . . . . . . . Better than ±0.1%

CO2 Range. . . . . . . . . . . . . 0-20%

Calibration. . . . . . . . . . . . . Auto-zero

 Inlet Pressure. . . . . . . . . . . 10 PSIG (0.7 bar)

Filter . . . . . . . . . . . . . . . . . 0.2 micron, disposable

Sensor . . . . . . . . . . . . . . . . IR

Controller . . . . . . . . . . . . . Microprocessor

Setpoint . . . . . . . . . . . . . . . Digital

Display . . . . . . . . . . . . . . . Digital LCD

Readability . . . . . . . . . . . . 0.1%

Setability . . . . . . . . . . . . . . 0.1%

Alarm Differential . . . . . . . ±1.0% (nominal) for approximately and Delay 30 minutes

Controlled RH

RH . . . . . . . . . . . . . . . . . . Ambient to >95% RH @37C (98.6F)

Humidity Control . . . . . . . ±2.0%

Sensor . . . . . . . . . . . . . . . . Platinum

Controller . . . . . . . . . . . . . Microprocessor

Setpoint . . . . . . . . . . . . . . . Digital

Display . . . . . . . . . . . . . . . Digital LCD

 Readability . . . . . . . . . . . . 0.1%

Setability . . . . . . . . . . . . . . 0.1%

Humidity Reservoir . . . . . . 1.0 gal. (3.8 liters)