Tuttnauer 3870E - B/L Steam Sterilizer Autoclave

  • 1.6 gal reservoir
  • 85L chamber
  • 3 prest programs
  • Automatic shut-off
  • 90-day warranty

The Tuttnauer 3870E – B/L autoclave is designed as a gravity displacement system for the sterilization of wrapped and unwrapped instruments and related items found in dental, medical and veterinary clinics, first aid rooms, hospitals, laboratories, etc. This autoclave is an electrically heated sterilizer using steam as the sterilizing agent.

A computerized control unit ensuring a fully automatic sterilization cycle, control and monitoring of physical parameters and a clear documentation of the sterilization cycle controls the autoclave. The 3870– B/L has three automatic programs, according to the material to be sterilized, and one auxiliary drying program.

CLICK HERE to see the Tuttnauer 3870E on our YouTube channel

A water pump is installed between the water reservoir and the chamber. This pump guarantees fast and accurate filling of the chamber every time. Entry of water may be accompanied by a noise for approximately 30 seconds. This is normal noise generated by regular operation of the pump. The control system provides adequate protection to ensure the safety of personnel and reliable operation with a minimum of shutdown time.

All of our equipment is used and reconditioned. Our technicians test all equipment to confirm that it operates properly and to the manufacturer’s specifications. While we can’t test for a specific analytical application, we will do our best to simulate the conditions under which the equipment will be run. If you have particular testing requirements or requests please let us know and we’ll work with you to make sure these requirements are addressed during testing.

The photos above are representative. Please bear in mind that the equipment shown is currently in storage in our warehouse and that in addition to testing each machine we will clean all the equipment thoroughly prior to shipment. We often have multiple units of each model in stock and sometimes we're unable to feature photos of every single unit. At your request, we can provide photos of the equipment in your order after it has been tested and cleaned.

Tuttnauer 3870E - B/L Steam Sterilizer Autoclave - Specifications

External dimensions 26.0”W x 34.5”D x 20.7”H
Chamber dimensions 15”dia x 30”L
Tray dimensions 11 (and 14)”W x 20”D x 1”H
Weight 278 lbs
Electrical requirements 230/240V, 50-60Hz, 13A

 