Fisher Isotemp 5150 Ultra-Low Temperature Chest Freezer

For some laboratories, protecting valuable biological samples is more important than maximizing storage capacity. When every degree matters and sample integrity takes priority, many researchers continue to choose chest-style ultralow freezers because they naturally retain cold air during routine access. The Fisher Isotemp 5150 Ultra-Low Temperature Chest Freezer was designed with that philosophy in mind, providing dependable storage from -50°C to -86°C in a compact chest configuration that minimizes temperature disturbance whenever the lid is opened.

With 3 cubic feet (85 liters) of storage capacity, the Fisher Isotemp 5150 is an excellent solution for laboratories requiring dedicated ultra-low storage for high-value samples, research collections, quality control materials, and specialty applications. Its compact footprint makes it particularly well suited for individual laboratories, hospital departments, university research groups, and facilities that need a secondary ultralow freezer close to the point of use.

Laboratories building a complete cold storage workflow often pair compact ultralow freezers with laboratory refrigerators for daily reagent storage, laboratory freezers for intermediate-temperature materials, and larger upright ultralow freezers for high-volume sample archives. The Isotemp 5150 serves a different purpose by providing exceptional thermal stability for smaller collections of critical samples.

Fisher Isotemp 5150 Ultra-Low Temperature Chest Freezer Key Features

  • 3 cu. ft. (85 L) storage capacity

  • Adjustable operating temperature from -50°C to -86°C

  • Chest-style cabinet minimizes cold air loss

  • Two-stage cascade refrigeration system

  • Microprocessor temperature controller

  • Digital control panel with 12-character message display

  • Audible and visual alarms

  • RS-232 and 4–20 mA outputs for external monitoring

  • Manual defrost operation

  • Refurbished and performance tested by American Instrument Exchange

  • 90-day warranty included

Fisher Isotemp 5150 Ultra-Low Temperature Chest Freezer Overview

Unlike upright freezers, chest ultralow freezers naturally retain cold air when opened because dense cold air remains inside the cabinet rather than spilling onto the laboratory floor. For laboratories retrieving samples throughout the day, this design helps minimize temperature fluctuations and supports rapid recovery following routine access.

The Fisher Isotemp 5150 combines this inherent thermal advantage with a dependable two-stage cascade refrigeration system, microprocessor controls, and comprehensive alarm monitoring. Its compact footprint allows laboratories to dedicate ultra-low storage to individual projects or research groups while maintaining dependable long-term sample preservation.

Many laboratories use the Isotemp 5150 alongside biosafety cabinets, CO₂ incubators, and refrigerated centrifuges, creating an efficient workflow from sample preparation through long-term preservation.

Common Applications

The Fisher Isotemp 5150 is commonly used for storing:

  • Cell lines

  • Stem cells

  • DNA and RNA libraries

  • Tissue specimens

  • Plasma and serum

  • Clinical trial samples

  • Protein samples

  • Enzymes

  • Vaccines

  • Pharmaceutical compounds

  • Microbial cultures

  • Biological reference materials

Its compact capacity makes it an excellent choice for laboratories requiring dedicated storage for a single research project, investigator, or specialty application where frequent sample access and maximum thermal stability are important.

Ideal Laboratory Types

The Isotemp 5150 is well suited for:

  • Academic research laboratories

  • Biotechnology companies

  • Pharmaceutical manufacturers

  • Hospital laboratories

  • Clinical research organizations

  • University research groups

  • Public health laboratories

  • Quality control laboratories

Its combination of compact size and chest-style thermal performance makes it especially attractive for laboratories with limited space or specialized storage requirements.

Temperature Recovery & Sample Protection

One of the greatest advantages of a chest ultralow freezer is its ability to retain cold air during routine access. Because cold air naturally settles inside the cabinet, opening the lid introduces less temperature disturbance than opening a conventional upright freezer. This design helps reduce unnecessary warming of stored samples while supporting faster recovery after lid openings.

The Fisher Isotemp 5150 combines this natural advantage with dependable cascade refrigeration, insulated cabinet construction, and programmable alarm monitoring to provide consistent long-term storage for valuable biological materials. Laboratories preserving especially sensitive collections frequently complement chest ultralow freezers with cryogenic storage systems for applications requiring liquid nitrogen preservation.

Typical Stored Materials

The Isotemp 5150 is commonly used to store:

  • DNA and RNA libraries

  • Cell lines

  • Stem cells

  • Tissue specimens

  • Plasma and serum

  • Vaccines

  • Protein samples

  • Enzymes

  • Pharmaceutical compounds

  • Clinical trial materials

  • Biological reference materials

Fisher Isotemp 5150 Ultra-Low Temperature Chest Freezer Key Specifications

Capacity

  • 3 cu. ft. (85 L)

Temperature Range

  • Adjustable from -50°C to -86°C

Refrigeration System

  • Two-stage cascade refrigeration

  • Manual defrost

Monitoring & Controls

  • Microprocessor controller

  • Digital display

  • 12-character message center

  • Audible and visual alarms

  • RS-232 communication

  • 4–20 mA monitoring output

Electrical Requirements

  • 120V, 60 Hz

Dimensions

  • Exterior: 28.7" W × 32.3" D × 43.8" H

  • Interior: 18.5" W × 18.5" D × 16.5" H

Weight

  • Approximately 432 lbs.

Energy Efficiency & Operating Costs

While chest ultralow freezers are often selected for their thermal performance rather than their energy consumption, the cabinet design can reduce temperature loss during routine access, decreasing refrigeration demand following lid openings. Combined with routine preventative maintenance and condenser cleaning, the Isotemp 5150 provides dependable long-term operation for laboratories prioritizing sample protection over storage volume.

Specifications compiled from Fisher Scientific manufacturer literature and American Instrument Exchange product information.

Refurbishment & Testing Process

Every Fisher Isotemp 5150 received by American Instrument Exchange undergoes a comprehensive inspection and extended performance evaluation before shipment. Our technicians verify pull-down performance, confirm stable operation throughout the operating temperature range, inspect the cascade refrigeration system and compressors, test alarms and electronic controls, verify digital temperature accuracy, inspect lid seals and insulation, and thoroughly clean the freezer before delivery.

Shipping & Installation Considerations

The Fisher Isotemp 5150 ships by freight from American Instrument Exchange. Before delivery, laboratories should verify floor space, electrical service, and adequate ventilation around the cabinet. Because chest freezers require overhead access, laboratories should also ensure sufficient clearance above the unit for comfortable daily operation.

Frequently Asked Questions

Why choose a chest ultralow freezer instead of an upright model?

Chest freezers naturally retain cold air when the lid is opened, helping minimize temperature fluctuations during routine sample retrieval. Many laboratories choose chest-style ULTs when protecting especially valuable samples or when the freezer will be accessed frequently.

What types of samples are best suited for the Isotemp 5150?

The Isotemp 5150 is commonly used for DNA and RNA libraries, stem cells, cell lines, tissue specimens, vaccines, plasma, serum, protein samples, pharmaceutical compounds, and other biological materials requiring storage as low as -86°C.

Can the Isotemp 5150 connect to external monitoring systems?

Yes. The freezer includes RS-232 and 4–20 mA outputs, allowing connection to independent monitoring and facility management systems for additional sample security.

How is the freezer tested before shipment?

Every refurbished Fisher Isotemp 5150 is inspected, cleaned, and performance tested by American Instrument Exchange technicians to verify pull-down performance, temperature stability, alarm functionality, refrigeration performance, and overall system operation before shipment.

CLICK HERE to see the Fisher Isotemp 5150 Ultralow Chest Freezer on our YouTube channel

Testing Procedures for Used and Refurbished Ultralow Freezers
Ensure -80°C Stability and Critical Sample Protection

Used and refurbished ultralow freezers from American Instrument Exchange are rigorously tested to ensure reliable long-term storage at temperatures as low as -80°C. Each unit undergoes a detailed inspection, full system evaluation, and extended runtime testing to meet the high standards required by research, clinical, and biorepository environments.

Why Testing Used Ultralow Freezers Is Essential

Before an ultralow freezer is placed back into service, it must be carefully tested for:
• Temperature stability and uniformity at ultralow levels (e.g., -70°C to -86°C)
• Cascade refrigeration system functionality and compressor health
• Seal and insulation effectiveness to prevent thermal loss
• Electrical safety, alarm functionality, and system compliance

Thorough testing helps prevent temperature excursions and equipment failure—crucial for protecting sensitive materials such as biological samples, vaccines, enzymes, and research specimens.
Our structured ULT testing protocol ensures each refurbished unit delivers dependable performance and long-term reliability.

Key Testing Procedures for Refurbished and Used Ultralow Freezers

1. Initial Visual Inspection

• Inspect external cabinet, hinges, and door gaskets for wear, corrosion, or damage
• Check power cords, plugs, and control interfaces for signs of wear or tampering
• Verify that internal racks, shelving, and chamber components are intact and operational

2. Functional Startup and Control Testing

• Power on the unit and confirm successful initialization of digital displays and alarm systems
• Test control panels, keypad responsiveness, and user-setpoint functionality
• Verify that door seals close tightly and form a secure barrier

3. Temperature and -80°C Stability Testing

• Test and evaluate internal temperature sensors thermometers/probes
• Monitor temperature performance at multiple chamber points over extended runtimes
• Confirm uniformity and stability at key setpoints, including -70°C, -80°C, and -86°C

4. Refrigeration System and Cascade Compressor Evaluation

• Test dual-stage (cascade) compressor function, startup times, and cycling behavior
• Evaluate evaporator coils, heat exchangers, and vacuum insulation panels for performance and integrity
• Check for refrigerant leaks, low-pressure faults, and abnormal compressor noise

5. Alarm and Safety System Verification

• Test high/low temperature alarms, power failure alerts, and battery backup systems (if equipped)
• Verify lockout controls, alarm delay settings, and door-open indicators

6. Final Quality Assurance and Documentation

• Run a continuous multi-day test cycle at -80°C to confirm sustained performance
• Document temperature logs, controller accuracy, and compressor behavior
• Complete a detailed service report

Dependable Performance from American Instrument Exchange

All refurbished ultralow freezers from American Instrument Exchange are tested to exacting standards for performance, safety, and reliability. With our rigorous inspection and testing procedures, you can trust your ULT freezer to deliver consistent -80°C storage from day one—ideal for research, clinical, pharmaceutical, and biobanking applications.

Repair and Refurbishment Protocol for Used Ultralow Freezers
-80°C Storage Performance from American Instrument Exchange

At American Instrument Exchange, every used ultralow temperature freezer (ULT freezer) undergoes a rigorous, multi-step refurbishment protocol designed to restore reliable -80°C performance. Whether upright, chest-style, or undercounter, our refurbished ULT freezers are tested and restored to deliver the cold storage precision required for biological samples, vaccines, enzymes, and other temperature-sensitive materials.

Why Choose a Refurbished Ultralow Freezer from American Instrument Exchange?

• Fully tested for -80°C cold storage stability and compressor performance
• Backed by a warranty
• A cost-effective alternative to expensive new ULT units
• Environmentally responsible through equipment reuse and waste reduction
• Serviced by experienced technicians specializing in laboratory-grade refrigeration

Step-by-Step Ultralow Freezer Refurbishment Protocol

Every ultralow freezer sold by American Instrument Exchange is carefully evaluated and restored through the following process:

1. Initial Intake & Inspection

• Exterior inspection for dents, corrosion, or physical damage
• Evaluation of door seals, hinges, and internal chamber components
• Documentation of missing parts or components requiring repair

2. Cleaning and Decontamination

• Comprehensive interior and exterior cleaning using non-abrasive cleaners
• Removal of dust, frost buildup, and residual contaminants
• Decontamination of all internal surfaces and removable parts like shelves and drawers

3. Mechanical and Electrical Repairs

• Replacement or repair of worn or damaged components (door latches, hinges, gaskets, etc.)
• Inspection and repair of control boards, displays, sensors, and switches
• Electrical system evaluation, including fuses, wiring, relays, and connectors

4. Ultralow Refrigeration System Assessment

• Testing of dual-stage compressors and cascade cooling system
• Evaluation of heat exchangers, evaporator coils, and vacuum insulation panels
• Inspection for refrigerant leaks, low pressure, or signs of compressor fatigue
• Verification of thermal insulation integrity

5. Temperature and Performance Testing

• Testing of internal temperature sensors and controllers
• Multi-point testing to verify chamber temperature uniformity at -80°C
• Long-duration, multi-day operational testing to ensure sustained ultralow performance

6. Safety and Compliance Checks

• Electrical safety testing and grounding verification
• Alarm function tests for high/low temperature alerts and power failure warnings

7. Final Quality Control and Documentation

• Multi-day runtime validation at -80°C
• Recording of temperature logs, compressor cycle data, and performance benchmarks
• Final inspection and service/testing report completed

Trust the Experts at American Instrument Exchange

With over 50 years in the lab equipment industry, American Instrument Exchange is your trusted source for refurbished ultralow freezers. Our certified refurbishment protocol ensures each ULT freezer delivers the precision, performance, and reliability your lab needs—at a fraction of the cost of new equipment.

Warranty Information

All used equipment sold by American Instrument Exchange is refurbished/reconditioned, tested to meet specifications, cleaned, and comes with a 90-day warranty from the date of shipment unless otherwise specified. New products include a warranty from the manufacturer. Your warranty is only valid if you have paid your invoice. If your invoice is over 30 days old, warranty work will not be performed until your invoice is paid in full.  American Instrument Exchange is not liable for any special or consequential damages that result from the use of, or the inability to use, the materials on this site or the performance of the products,

Exceptions include international sales and items that are agreed to be sold with a lesser warranty or items that are agreed to be sold “as is.”

Should there be visible shipping damage to packaging or concealed shipping damage to item(s) supplied, buyer must communicate such damage to the delivering entity or carrier upon receipt first and to American Instrument the same day item is accepted / signed for. Failure to do so voids warranty. Customer decision to involve / procure third party or OEM service other than American Instrument or AIE assigned vendors to resolve in-warranty instrument failure or defect without first placing a service call to American Instrument voids warranty.

Return & Cancellation Policy

American Instrument Exchange will authorize the return of and issue a full refund for any equipment that is not working as represented and could not be replaced or repaired to the customer’s satisfaction by AIE or a representative of AIE during the warranty period. Returns without prior authorization and an RMA number will not be accepted. 

Customers are responsible for proper packaging of the equipment and all included accessories. American Instrument will pay for insurance and shipping costs. Non-inventory items bought specifically to fill customer orders may possibly be returned, but on a case by case basis.

Customers who cancel orders after refurbishing has begun or after delivery where there are no equipment performance issues may be subject to a Restocking/Cancellation Fee of 20% or more and responsible for shipping costs for both the delivery and return.

Ongoing and Extended Support Options

American Instrument Exchange is able to provide additional repair and support services outside the warranty terms for most products that sell.

For biosafety cabinets, clean benches, chemical fume hoods, and other clean air products our team at AIE Laboratory Services can provide certifications, repairs, decontaminations, and other services for customers throughout New England, and guidance and limited technical assistance for any customers in the US and Canada for new or refurbished biosafety cabinets purchased from American Instrument Exchange.

For all other products, the repair service team at American Instrument Exchange can provide both field and depot repair evaluations. We can easily schedule and coordinate the pick-up and transportation of your malfunctioning equipment. By offering hourly rates lower than most OEM service providers and eliminating travel time for technicians, we help reduce repair costs. Furthermore, our lab depot repair service minimizes or eliminates scheduling delays. Our experienced service technicians are qualified to restore your equipment to factory specifications.

Please contact our team at info@americaninstrument.com for more details.

FAQ

Does the equipment come with a warranty?

Yes, all equipment from American Instrument Exchange comes with a warranty.
Used/reconditioned/refurbished equipment is tested to make sure it operates to the manufacturer's specifications and comes with a 90-day warranty (unless otherwise noted). New equipment comes with a warranty from the manufacturer.

How do you test and recondition the equipment?

Testing requirements vary across all the equipment categories we sell. Temperature controlled units are tested with NIST traceable probes and data loggers. Gaskets, external housing, glass, handles, controls and control screens are all thoroughly inspected for integrity and functionality.
We test all equipment using the manufacturer's parameters and specifications to confirm that the unit operates properly.

How do products ship from American Instrument Exchange?

We can ship products both nationally and internationally using a well-established network of transporters. Smaller items will ship via UPS. Larger equipment will be palletized and shipped either LTL or air-ride, depending on the specific equipment and/or delivery requirements.

Return and order cancellation policy

American Instrument Exchange will accept the return of any product that is not working as represented and could not be replaced or repaired to the customer's satisfaction during the warranty period. Customers who cancel orders after refurbishing has begun will be subject to a cancellation and restocking fee of between 20% and 40%.

Are discounts available for educational or non-profit institutions, or for bulk purchases?

Yes – call or email us today with a list of equipment you're looking for and we can potentially put together a quote that includes a discount.

How does your price matching policy work?

We strive to keep our pricing competitive and reasonable. If you find comparable items for sale at a lower price from other vendors please send us a link and or quote for the item and we'll do our best to match or beat the price.

Can I see the equipment in person before committing to a purchase?

Customers are welcome to make an appointment to view equipment in our warehouse. Given our workload we may not be able to test or clean the equipment prior to your arrival if you have not committed to an order.

1023 Western Avenue is a working warehouse, not a showroom, and customers should dress appropriately and wear closed toe shoes.

Can I ship products that I purchased elsewhere to American Instrument for repair?

Yes – please see our repair services page for more information.