Forma 3033 CO2 SteriCult 200
Contact Us
Call 978-521-2221
to request information & pricing for item 3172 INC
American Instrument Exchange
Contact us for all your used lab equipment needs
1023 Western Ave.
Haverhill, MA 01832
Phone: 978-521-2221
Fax: 978-521-8822
Email: info@americaninstrument.com
Shipping Policy
Thermo Stericult 3033
- Single 10.5 cubic foot chamber
- Temperature range: ambient +4°C to 45°C
- CO2 range: 0 – 20%
- 90 day warranty
The Thermo Forma Steri-Cult CO2 3033 Incubator uses HEPA filtered, vertical laminar airflow techniques and a filtered external water supply to minimize contamination and desiccation of cultures. The Thermo Forma 3033 has an innovative humidity supply system located outside the chamber that provides rapid RH recovery. The humidity system is not a vapor generator and has no immersible heater to burn out or to cause scaling. Water reservoir for the humidity system is easily filled by way of a port on the outside of the unit. Humidity control is user-selectable above ambient for total control over chamber conditions.
Most of our equipment is used and reconditioned. We test CO2 incubators using a calibrated digital chart recorder to verify the accuracy of temperature controls and readouts, as well as using a Fyrite indicator to measure CO2 levels in the chamber. We replace any hardware or parts as is necessary and clean all incubators thoroughly prior to shipment.
SPECIFICATIONS
Temperature
Control . . . . . . . . . . . . . . . ±0.1C
Range . . . . . . . . . . . . . . . . 4C above ambient to 45C (113F)
Sensor . . . . . . . . . . . . . . . . Platinum
Controller . . . . . . . . . . . . . Microprocessor
Setpoint . . . . . . . . . . . . . . . Digital
Display . . . . . . . . . . . . . . . Digital LCD
Readability . . . . . . . . . . . . 0.1C
Setability . . . . . . . . . . . . . . 0.1C
Uniformity. . . . . . . . . . . . . ±0.3C @ 37C (98.6F)
Temperature Alarm
Sensor . . . . . . . . . . . . . . . . Platinum
Controller . . . . . . . . . . . . . Independent analog (high and low alarm)
Setpoint . . . . . . . . . . . . . . . Digital with screwdriver adjust
Display . . . . . . . . . . . . . . . Audible/visual
Readability . . . . . . . . . . . . 0.1C
Setability . . . . . . . . . . . . . . 0.1C
CO2
Control . . . . . . . . . . . . . . . Better than ±0.1%
CO2 Range. . . . . . . . . . . . . 0-20%
Calibration. . . . . . . . . . . . . Auto-zero
Inlet Pressure. . . . . . . . . . . 10 PSIG (0.7 bar)
Filter . . . . . . . . . . . . . . . . . 0.2 micron, disposable
Sensor . . . . . . . . . . . . . . . . IR
Controller . . . . . . . . . . . . . Microprocessor
Setpoint . . . . . . . . . . . . . . . Digital
Display . . . . . . . . . . . . . . . Digital LCD
Readability . . . . . . . . . . . . 0.1%
Setability . . . . . . . . . . . . . . 0.1%
Alarm Differential . . . . . . . ±1.0% (nominal) for approximately and Delay 30 minutes
Controlled RH
RH . . . . . . . . . . . . . . . . . . Ambient to >95% RH @37C (98.6F)
Humidity Control . . . . . . . ±2.0%
Sensor . . . . . . . . . . . . . . . . Platinum
Controller . . . . . . . . . . . . . Microprocessor
Setpoint . . . . . . . . . . . . . . . Digital
Display . . . . . . . . . . . . . . . Digital LCD
Readability . . . . . . . . . . . . 0.1%
Setability . . . . . . . . . . . . . . 0.1%
Humidity Reservoir . . . . . . 1.0 gal. (3.8 liters)