Brand : Mettler Toledo
Product Code : SKU 0665KK BALANCE

Shipping

All products ship from our 30,000 sq. ft. Haverhill, MA warehouse

Mettler Toledo XS105 Analytical Balance – Used & Refurbished

The Mettler Toledo XS105 Analytical Balance is a high-resolution laboratory balance designed for demanding quantitative applications requiring substantially greater sensitivity than a conventional 0.1 mg analytical balance. With a 120 g maximum capacity and 0.01 mg (10 µg) readability, the XS105 is well suited for pharmaceutical research, analytical chemistry, standards preparation, formulation, gravimetric analysis, quality control, and other procedures involving small samples and precise mass determination.

Available used and refurbished from American Instrument Exchange, the Mettler Toledo XS105 provides laboratories with proven Excellence-series weighing technology at a significantly lower acquisition cost than comparable new high-resolution balances. Each XS105 is inspected, cleaned, calibrated, and tested for weighing accuracy and repeatability before shipment and includes a 90-day warranty from American Instrument Exchange.

Key Features

  • 120 g maximum weighing capacity

  • 0.01 mg / 10 µg readability

  • Typical repeatability of approximately 0.008 mg / 8 µg

  • Typical USP minimum weight of approximately 16 mg

  • ±0.2 mg linearity

  • Approximately 3-second settling time

  • FACT automatic internal adjustment

  • SmartGrid weighing pan minimizes turbulence inside the weighing chamber

  • Enclosed draft shield protects high-resolution measurements from air movement

  • ErgoClip compatibility for secure placement of different weighing vessels

  • Detachable terminal allows flexible workstation configuration

  • Tool-free removable draft shield components simplify cleaning

  • LabX software compatibility

  • Used, refurbished, calibrated, and tested by American Instrument Exchange

  • 90-day warranty included

Overview

The Mettler Toledo XS105 is not a conventional 0.1 mg analytical balance. Its 0.01 mg, or 10 µg, readability provides ten times finer resolution, placing the XS105 in the high-resolution analytical range between standard analytical balances and dedicated microbalances.

That combination of resolution and 120 g capacity makes the XS105 particularly useful for laboratories working with small quantities but still requiring enough capacity for conventional weighing vessels, containers, filters, and other analytical samples.

Mettler Toledo designed the XS series to address some of the practical difficulties associated with high-resolution weighing. Its SmartGrid weighing pan minimizes turbulence within the weighing chamber, helping measurements stabilize more rapidly. The enclosed draft shield protects the weighing system from room air currents, while the open-grid pan design reduces the amount of solid surface area exposed to moving air.

The balance also incorporates FACT automatic internal adjustment, helping maintain weighing performance as environmental conditions change.

For laboratories that need greater sensitivity than a standard analytical balance but do not require the 1 µg readability of a dedicated microbalance, the XS105 offers a particularly useful combination of 10 µg resolution, 120 g capacity, ergonomic sample handling, and proven Mettler Toledo weighing technology.

Common Applications

The Mettler Toledo XS105 is particularly well suited for applications involving small samples and high-resolution quantitative measurements.

Pharmaceutical and biotechnology laboratories can use the XS105 for standards preparation, formulation research, active ingredient weighing, reagent preparation, sample preparation, quality control, and other procedures where small differences in mass must be measured reliably.

Analytical chemistry laboratories may use the XS105 for gravimetric analysis, solution preparation, reference-standard preparation, differential weighing, density determination, filter weighing, and other high-resolution quantitative procedures.

Environmental laboratories can use the balance for filter and particulate measurements, while materials science and academic research laboratories may use it for small-sample preparation, comparative weighing, material characterization, and experimental work.

Its combination of 10 µg readability and a relatively generous 120 g capacity also makes the XS105 useful when laboratories need high resolution without the significantly smaller capacities typically associated with true microbalances.

Ideal Laboratory Types

The XS105 is a strong choice for pharmaceutical, biotechnology, analytical chemistry, environmental testing, materials science, academic research, formulation, manufacturing, and quality-control laboratories.

It is particularly appropriate for laboratories that have reached the practical limits of conventional 0.1 mg analytical balances.

For organizations routinely weighing small samples or preparing low-concentration standards, the tenfold improvement in displayed resolution can provide a meaningful advantage. At the same time, the 120 g maximum capacity gives operators considerably more flexibility than many dedicated microbalances.

This makes the XS105 useful as either a primary high-resolution analytical balance or as a specialized instrument alongside standard analytical and precision balances.

High-Resolution Analytical Weighing Performance

The defining feature of the Mettler Toledo XS105 is its 0.01 mg, or 10 µg, readability.

Mettler Toledo specifies typical repeatability of approximately 0.008 mg, or 8 µg, along with linearity of ±0.2 mg. Typical settling time is approximately 3 seconds.

These specifications position the XS105 substantially above a conventional analytical balance in terms of resolution while retaining a useful 120 g weighing capacity.

High-resolution performance also means that environmental influences become increasingly important. At the 10 µg level, small disturbances that may be insignificant on a precision balance can become visible in the weighing result.

Mettler Toledo's SmartGrid weighing pan helps address this problem by reducing turbulence within the weighing chamber. Compared with a conventional solid weighing pan, the open-grid design presents less surface area to moving air and allows measurements to stabilize efficiently inside the enclosed draft shield.

For laboratories performing repeated high-resolution measurements, this combination of sensitivity and stabilization can improve both measurement quality and workflow efficiency.

Minimum Sample Weight & USP Applications

Minimum sample weight is a critical consideration when a balance is used for pharmaceutical or other regulated quantitative procedures.

Mettler Toledo specifies a typical USP minimum weight of approximately 16 mg for the XS105 under stated conditions.

This relatively low typical minimum weight reflects the balance's high-resolution weighing performance and makes the XS105 potentially useful for applications involving substantially smaller sample quantities than can generally be handled reliably on a standard 0.1 mg analytical balance.

Actual minimum weight is not determined by readability alone. It depends on the repeatability of the individual installed balance, environmental conditions, installation quality, measurement uncertainty requirements, and applicable laboratory procedures.

For regulated applications, laboratories should establish and document minimum weight according to their own balance qualification procedures and quality system rather than relying exclusively on a manufacturer's typical published value.

Controlling Drafts, Vibration & Static

At 10 µg readability, environmental control becomes an important part of reliable weighing.

Air currents, vibration, temperature gradients, static electricity, unstable work surfaces, and operator technique can all influence high-resolution measurements.

The XS105 uses an enclosed draft shield to isolate the weighing chamber from room air movement. Its SmartGrid weighing pan further reduces turbulence around the sample and helps the weighing system stabilize efficiently.

Static electricity deserves particular attention when weighing powders, filters, plastics, glass vessels, and other materials capable of accumulating electrostatic charge. Electrostatic forces can produce apparent changes in mass that are significant relative to the XS105's 10 µg readability.

Mettler Toledo offered antistatic accessories for the XS series to help neutralize charged samples and containers when static becomes a significant source of measurement error.

For best performance, the XS105 should be installed on a rigid, level, vibration-resistant surface away from HVAC vents, doors, direct sunlight, centrifuges, shakers, refrigeration compressors, and other sources of vibration or temperature fluctuation.

A dedicated balance table or anti-vibration workstation may be beneficial when the instrument is routinely used near the lower end of its weighing range.

Calibration & Internal Adjustment

The XS105 incorporates Mettler Toledo's FACT automatic internal adjustment system.

FACT helps compensate for changes in operating conditions by allowing the balance to perform internal adjustment automatically. This reduces reliance on manually initiated adjustment procedures and helps maintain consistent weighing performance throughout routine laboratory operation.

Automatic internal adjustment does not eliminate the need for external performance verification.

Laboratories operating under GLP, GMP, USP, or internal quality systems may require scheduled testing using appropriate traceable calibration weights along with documented accuracy, repeatability, and minimum-weight verification.

American Instrument Exchange evaluates refurbished balances using appropriate calibration weights and verifies critical weighing characteristics before shipment.

SmartGrid, ErgoClips & Sample Handling

High-resolution weighing is influenced not only by the balance itself but also by how the sample and weighing vessel interact with the weighing chamber.

The XS105 uses Mettler Toledo's distinctive SmartGrid weighing pan, designed to reduce the effects of air turbulence and improve stabilization.

Mettler Toledo's ErgoClip accessories can be attached to the SmartGrid to securely support different types of weighing vessels. This allows samples to be weighed directly into appropriate containers while minimizing awkward positioning and unnecessary sample transfers.

Direct weighing into the final container can be particularly valuable when handling expensive compounds, small sample quantities, or materials where transfer losses could significantly affect the final result.

The draft shield can also be disassembled without tools, simplifying access to the weighing chamber for routine cleaning. This is especially useful in pharmaceutical, analytical, and research laboratories where powders and other small particles may accumulate around the weighing area.

Data Management & LabX Compatibility

The XS105 can be integrated into more structured laboratory data workflows through compatibility with Mettler Toledo LabX software.

LabX can support automatic calculations, data storage, reporting, and standardized weighing procedures, helping reduce reliance on manual transcription.

This capability can be especially useful in pharmaceutical, biotechnology, and quality-control laboratories where weighing results become part of formal analytical records.

Laboratories evaluating a refurbished XS105 for integration with existing data systems should confirm software versions, interfaces, licensing, and compatibility with their current laboratory infrastructure.

Key Specifications

Specification Mettler Toledo XS105
Balance Type High-Resolution Analytical Balance
Maximum Capacity 120 g
Readability 0.01 mg / 10 µg
Typical Repeatability 0.008 mg / 8 µg
Typical USP Minimum Weight 16 mg
Linearity ±0.2 mg
Settling Time Approx. 3 seconds
Adjustment Internal / FACT
Weighing Pan 73 × 78 mm
Weighing Pan Technology SmartGrid
Draft Protection Enclosed draft shield
Sample Holders ErgoClip compatible
Display Detachable terminal
Overall Dimensions 263 mm W × 453 mm D × 322 mm H
Software Compatibility LabX
Legal for Trade No

Choosing Between an Analytical, Semi-Micro & Microbalance

Choosing between an analytical, semi-micro, and microbalance requires considering more than maximum capacity. The smallest quantity that must be measured reliably is often the more important factor.

A conventional analytical balance commonly provides 0.1 mg readability and is appropriate for a broad range of routine quantitative laboratory procedures.

The XS105 provides 0.01 mg, or 10 µg, readability, giving it ten times finer displayed resolution. This makes it a strong choice for laboratories working with smaller quantities or requiring greater sensitivity than a standard analytical balance can provide.

A dedicated microbalance may provide 1 µg readability, offering another tenfold increase in resolution. That additional sensitivity, however, generally comes with lower capacity, greater environmental sensitivity, and more demanding installation requirements.

The XS105 therefore occupies a valuable middle ground. Laboratories receive near-semi-micro/high-resolution analytical performance while retaining 120 g of weighing capacity.

For extremely small samples where true microgram-level readability is required, a microbalance such as the Mettler Toledo XPE26 may be more appropriate. For routine measurements where 0.1 mg readability is sufficient, a conventional analytical balance may offer a more economical solution.

Refurbishment & Performance Testing

Every used Mettler Toledo XS105 sold by American Instrument Exchange is inspected, cleaned, refurbished as necessary, calibrated, and tested before shipment.

High-resolution analytical balances require substantially more evaluation than a simple power-on test. Our technicians use appropriate calibration weights to check critical performance characteristics including accuracy, repeatability, linearity, corner-load tolerances, stability, and reduced span at or near full capacity.

The FACT internal adjustment system, weighing cell, display, controls, SmartGrid pan, draft shield, leveling components, and other installed features are also inspected and tested for proper operation.

The weighing chamber, SmartGrid, draft shield panels, and other accessible components are thoroughly cleaned before shipment.

Each refurbished XS105 includes a 90-day warranty from American Instrument Exchange.

Shipping & Installation Considerations

A balance capable of resolving 10 µg changes in mass requires careful handling during transportation and thoughtful installation after delivery.

American Instrument Exchange securely prepares the XS105 for shipment to protect the weighing system, SmartGrid pan, draft shield, display, and other components.

Once received, the balance should be installed on a rigid, level, vibration-resistant work surface in an area with stable temperature and minimal air movement. The instrument should be allowed to acclimate to its new laboratory environment before final adjustment, calibration verification, and critical weighing procedures are performed.

For demanding high-resolution applications, a dedicated balance table or anti-vibration workstation should be considered.

Laboratories performing regulated measurements should follow installation with appropriate calibration, verification, minimum-weight determination, or qualification according to their quality system.

Related Laboratory Equipment

American Instrument Exchange maintains an extensive inventory of used and refurbished laboratory balances, including analytical balances, semi-micro balances, microbalances, and precision balances for applications ranging from routine sample preparation to high-resolution quantitative analysis.

Labs developing complete analytical or sample-preparation workflows can also source centrifuges, water baths, incubators, laboratory refrigerators, laboratory freezers, and other refurbished laboratory equipment from American Instrument Exchange.

Mettler Toledo XS105 Analytical Balance FAQs

What is the maximum capacity of the Mettler Toledo XS105?

The XS105 has a maximum weighing capacity of 120 g, providing useful capacity for high-resolution analytical weighing applications.

What is the readability of the XS105?

The Mettler Toledo XS105 provides 0.01 mg, or 10 µg, readability. This is ten times finer than the 0.1 mg readability commonly associated with conventional analytical balances.

What is the typical repeatability of the XS105?

Mettler Toledo specifies typical repeatability of approximately 0.008 mg, or 8 µg, under appropriate operating conditions.

What is the typical USP minimum weight of the XS105?

Mettler Toledo specifies a typical USP minimum weight of approximately 16 mg under stated conditions. Actual minimum weight should be determined according to the installed balance's repeatability, environmental conditions, measurement uncertainty, and laboratory quality requirements.

Is the XS105 an analytical balance or a microbalance?

The XS105 is a high-resolution analytical balance. Its 10 µg readability provides substantially greater resolution than a conventional 0.1 mg analytical balance but does not reach the 1 µg readability typically associated with a dedicated microbalance.

What is SmartGrid on the Mettler Toledo XS105?

SmartGrid is an open-grid weighing pan designed to reduce turbulence within the weighing chamber. By minimizing the effects of moving air around the sample, SmartGrid helps high-resolution measurements stabilize efficiently.

What are ErgoClips?

Mettler Toledo ErgoClips are accessories designed to attach to the SmartGrid weighing pan and securely hold different types of weighing vessels. They can allow samples to be weighed directly into their intended containers, reducing sample transfers and potential material loss.

Does the XS105 have automatic internal calibration?

Yes. The XS105 incorporates FACT automatic internal adjustment, helping maintain weighing performance as environmental conditions change. Laboratories operating under regulated quality systems may still require documented external calibration and performance verification.

Is the XS105 compatible with LabX?

Yes. The XS105 is compatible with Mettler Toledo LabX software, allowing weighing results to be incorporated into automated calculation, data-storage, reporting, and standardized workflow processes when appropriately configured.

Why purchase a refurbished Mettler Toledo XS105?

A refurbished XS105 provides laboratories with 10 µg readability, 120 g capacity, SmartGrid weighing technology, FACT internal adjustment, and LabX compatibility without the acquisition cost associated with comparable new high-resolution analytical balances. American Instrument Exchange tests, calibrates, cleans, and refurbishes each balance before shipment and provides a 90-day warranty.

Testing Procedures for Used and Refurbished Laboratory Balances

Ensure Precision, Repeatability, and Weighing Accuracy

Used and refurbished laboratory balances from American Instrument Exchange undergo thorough, multi-point testing to ensure accurate, reliable performance in scientific, industrial, and pharmaceutical settings. Each balance is carefully inspected, calibrated, and evaluated to meet the stringent standards required for analytical, precision, and micro weighing applications.

Why Testing Used Laboratory Balances Is Essential

Before a laboratory balance is placed back into service, it must be rigorously tested for:
• Weighing accuracy across full capacity range
• Repeatability and linearity
• Load cell and sensor functionality
• User interface and calibration system integrity

Comprehensive balance testing helps eliminate measurement errors and guarantees performance in high-precision environments such as R&D, formulation labs, and quality control facilities.

Our structured balance testing protocol ensures every refurbished unit delivers consistent results and trusted performance.

Key Testing Procedures for Refurbished and Used Laboratory Balances

1. Initial Visual Inspection

• Inspect housing, leveling feet, draft shield, and weighing pan for wear, cracks, or corrosion
• Check control panel, power cord, and buttons for damage or signs of tampering
• Confirm presence and condition of all required components (e.g., shields, adapters, trays)

2. Functional Startup and Interface Testing

• Power on the balance to verify display initialization and zeroing functionality
• Test user interface responsiveness, menu access, and programmable features
• Evaluate stabilization time and auto-tare functions

3. Calibration and Accuracy Testing

• Perform internal calibration (if equipped) and verify operation
• Use certified calibration weights to perform external calibration
• Check for accuracy at multiple weight points across the balance’s range

4. Repeatability, Linearity, and Corner Load Testing

• Conduct multiple weighings of identical test weights to confirm repeatability
• Test linearity by weighing at low, mid, and high points across the range
• Evaluate corner-load accuracy by placing test weights at different positions on the pan

5. Load Cell and Sensor Integrity Verification

• Test load cell response for consistency, drift, or hysteresis
• Verify sensor functionality and precision at sub-milligram levels (for analytical/microbalances)

6. Electrical Safety and Grounding Checks

• Inspect wiring and power components for safety and compliance
• Verify proper grounding and electrical insulation to meet lab equipment standards

7. Final Quality Assurance and Documentation

• Perform a full operational test under simulated lab conditions
• Record and document all calibration results and test metrics
• Final cleaning and inspection before shipping

Trusted Performance from American Instrument Exchange

All refurbished laboratory balances from American Instrument Exchange are tested and validated to deliver precision weighing performance from day one. Our expert technicians use industry-standard procedures to ensure each balance is lab-ready—whether used for sample prep, formulation, or high-precision research.

Repair and Refurbishment Protocol for Used Laboratory Balances

Precision Weighing Restored by American Instrument Exchange

At American Instrument Exchange, every used laboratory balance is subjected to a multi-step refurbishment protocol designed to restore precision and reliability with laboratory standards. Whether it’s an analytical balance, precision balance, or microbalance, we ensure each unit delivers consistent performance for scientific, pharmaceutical, and industrial applications.

Why Choose a Refurbished Laboratory Balance from American Instrument Exchange?

• Fully tested for accuracy, repeatability, and calibration stability
• Backed by a warranty
• Cost-effective alternative to purchasing new balances
• Environmentally responsible through expert refurbishment and reuse
• Refurbished by technicians with extensive experience in lab instrumentation

Step-by-Step Laboratory Balance Refurbishment Protocol

Every laboratory balance from American Instrument Exchange is professionally restored through a comprehensive, multi-point process:

1. Initial Intake & Inspection

• Exterior inspection for damage, corrosion, and wear on display, housing, and leveling feet
• Visual/manual check of weighing pan, draft shield, doors, and buttons
• Documentation of missing parts, error codes, or component issues

2. Cleaning and Decontamination

• Thorough cleaning of the weighing chamber, pan, and internal components using anti-static and non-abrasive cleaners
• Removal of dust, chemical residue, and particulate buildup
• Disinfection of all surfaces

3. Mechanical and Electronic Repairs

• Replacement/repair of worn components including pans, draft shields, and buttons
• Evaluation and repair of load cells, sensors, and electronic boards
• Inspection of power supply, internal wiring, and display screen functionality

4. Calibration and Accuracy Testing

• Internal calibration mechanisms are tested (if equipped)
• External calibration is performed using certified traceable weights
• Repeatability, linearity, and corner-load accuracy are verified

5. Functional Testing and Performance Validation

• Test of user interface, menu settings, and programmable functions
• Taring, zeroing, and stability response times are evaluated
• Load-bearing accuracy is tested across the unit’s full capacity range

6. Compliance and Safety Checks

• Electrical system testing for grounding and safe operation
• Verification of leveling system and anti-vibration performance

7. Final Quality Control and Documentation

• Final inspection and multi-point accuracy verification
• Complete operational test with weight logs and calibration reports
• Full cleaning and packaging for shipment

Trust the Experts at American Instrument Exchange

With over 50 years of experience in laboratory equipment, American Instrument Exchange is a trusted source for precision-refurbished laboratory balances. Our refurbishment protocol ensures each balance meets strict standards for accuracy and performance—ready to deliver dependable weighing results from day one.

Warranty Information

All used equipment sold by American Instrument Exchange is refurbished/reconditioned, tested to meet specifications, cleaned, and comes with a 90-day warranty from the date of shipment unless otherwise specified. New products include a warranty from the manufacturer. Your warranty is only valid if you have paid your invoice. If your invoice is over 30 days old, warranty work will not be performed until your invoice is paid in full.  American Instrument Exchange is not liable for any special or consequential damages that result from the use of, or the inability to use, the materials on this site or the performance of the products,

Exceptions include international sales and items that are agreed to be sold with a lesser warranty or items that are agreed to be sold “as is.”

Should there be visible shipping damage to packaging or concealed shipping damage to item(s) supplied, buyer must communicate such damage to the delivering entity or carrier upon receipt first and to American Instrument the same day item is accepted / signed for. Failure to do so voids warranty. Customer decision to involve / procure third party or OEM service other than American Instrument or AIE assigned vendors to resolve in-warranty instrument failure or defect without first placing a service call to American Instrument voids warranty.

Return & Cancellation Policy

American Instrument Exchange will authorize the return of and issue a full refund for any equipment that is not working as represented and could not be replaced or repaired to the customer’s satisfaction by AIE or a representative of AIE during the warranty period. Returns without prior authorization and an RMA number will not be accepted. 

Customers are responsible for proper packaging of the equipment and all included accessories. American Instrument will pay for insurance and shipping costs. Non-inventory items bought specifically to fill customer orders may possibly be returned, but on a case by case basis.

Customers who cancel orders after refurbishing has begun or after delivery where there are no equipment performance issues may be subject to a Restocking/Cancellation Fee of 20% or more and responsible for shipping costs for both the delivery and return.

Ongoing and Extended Support Options

American Instrument Exchange is able to provide additional repair and support services outside the warranty terms for most products that sell.

For biosafety cabinets, clean benches, chemical fume hoods, and other clean air products our team at AIE Laboratory Services can provide certifications, repairs, decontaminations, and other services for customers throughout New England, and guidance and limited technical assistance for any customers in the US and Canada for new or refurbished biosafety cabinets purchased from American Instrument Exchange.

For all other products, the repair service team at American Instrument Exchange can provide both field and depot repair evaluations. We can easily schedule and coordinate the pick-up and transportation of your malfunctioning equipment. By offering hourly rates lower than most OEM service providers and eliminating travel time for technicians, we help reduce repair costs. Furthermore, our lab depot repair service minimizes or eliminates scheduling delays. Our experienced service technicians are qualified to restore your equipment to factory specifications.

Please contact our team at info@americaninstrument.com for more details.

FAQ

Does the equipment come with a warranty?

Yes, all equipment from American Instrument Exchange comes with a warranty.
Used/reconditioned/refurbished equipment is tested to make sure it operates to the manufacturer's specifications and comes with a 90-day warranty (unless otherwise noted). New equipment comes with a warranty from the manufacturer.

How do you test and recondition the equipment?

Testing requirements vary across all the equipment categories we sell. Temperature controlled units are tested with NIST traceable probes and data loggers. Gaskets, external housing, glass, handles, controls and control screens are all thoroughly inspected for integrity and functionality.
We test all equipment using the manufacturer's parameters and specifications to confirm that the unit operates properly.

How do products ship from American Instrument Exchange?

We can ship products both nationally and internationally using a well-established network of transporters. Smaller items will ship via UPS. Larger equipment will be palletized and shipped either LTL or air-ride, depending on the specific equipment and/or delivery requirements.

Return and order cancellation policy

American Instrument Exchange will accept the return of any product that is not working as represented and could not be replaced or repaired to the customer's satisfaction during the warranty period. Customers who cancel orders after refurbishing has begun will be subject to a cancellation and restocking fee of between 20% and 40%.

Are discounts available for educational or non-profit institutions, or for bulk purchases?

Yes – call or email us today with a list of equipment you're looking for and we can potentially put together a quote that includes a discount.

How does your price matching policy work?

We strive to keep our pricing competitive and reasonable. If you find comparable items for sale at a lower price from other vendors please send us a link and or quote for the item and we'll do our best to match or beat the price.

Can I see the equipment in person before committing to a purchase?

Customers are welcome to make an appointment to view equipment in our warehouse. Given our workload we may not be able to test or clean the equipment prior to your arrival if you have not committed to an order.

1023 Western Avenue is a working warehouse, not a showroom, and customers should dress appropriately and wear closed toe shoes.

Can I ship products that I purchased elsewhere to American Instrument for repair?

Yes – please see our repair services page for more information.