VWR Explosion-Proof Laboratory Refrigerator

The VWR Explosion-Proof Laboratory Refrigerator is a laboratory-grade refrigerator engineered for the safe refrigerated storage of flammable and volatile chemicals in hazardous locations where ignitable vapors may be present. Offering 14 cubic feet (396.4 L) of refrigerated storage capacity and an operating temperature range of 0°C to 10°C, this explosion-proof refrigerator is designed for chemical laboratories, pharmaceutical manufacturers, petrochemical facilities, research institutions, and industrial laboratories requiring refrigeration equipment suitable for hazardous environments.

Unlike flammable material storage refrigerators, which are intended for use in non-hazardous laboratory spaces, the VWR Explosion-Proof Laboratory Refrigerator incorporates explosion-proof electrical construction and a hermetically sealed refrigeration system to reduce ignition hazards where combustible vapors may be present. Non-sparking magnetic door gaskets, heavy-duty cabinet construction, and hardwired installation further enhance laboratory safety while providing dependable refrigerated storage for hazardous materials. The refrigerator is designed for use in accordance with NFPA 45 and NFPA 70 guidelines for hazardous laboratory environments.

Many laboratories pair explosion-proof refrigeration with chemical fume hoods, biosafety cabinets, laboratory freezers, and centrifuges to create a complete hazardous-material handling workflow.

VWR Explosion-Proof Laboratory Refrigerator Key Features

  • Manufacturer: VWR

  • Product Type: Explosion-Proof Laboratory Refrigerator

  • 14 cu. ft. (396.4 L) storage capacity

  • Temperature range: 0°C to 10°C

  • Explosion-proof electrical construction

  • Hermetically sealed compressor

  • Non-sparking magnetic door gaskets

  • Heavy-duty painted steel cabinet

  • One adjustable and two fixed shelves

  • Hardwired installation (no line cord or plug)

  • Designed for hazardous laboratory locations

  • Meets NFPA 45 & NFPA 70 recommendations for hazardous-location refrigerated storage

  • Refurbished and performance tested by American Instrument Exchange

  • 90-day warranty included.

Overview

The VWR Explosion-Proof Laboratory Refrigerator is specifically designed for laboratories where flammable vapors may be present during normal operation. Unlike conventional laboratory refrigerators, all potential ignition sources are either removed, sealed, or enclosed within explosion-proof electrical components, allowing the refrigerator to operate safely in hazardous work areas.

The refrigerator utilizes a hermetically sealed compressor and non-sparking magnetic door gaskets to provide dependable temperature performance while minimizing ignition risks. Its durable laboratory construction and straightforward refrigeration system make it well suited for long-term service in demanding chemical laboratory environments.

Common Applications

The VWR Explosion-Proof Laboratory Refrigerator is commonly used for storing volatile solvents, chromatography reagents, alcohols, petroleum products, pharmaceutical ingredients, laboratory standards, chemical reference materials, volatile organic compounds (VOCs), and other hazardous chemicals requiring refrigerated storage. It is frequently installed in pharmaceutical manufacturing facilities, university chemistry departments, industrial quality-control laboratories, petrochemical facilities, and government research laboratories where hazardous atmospheres may exist.

Ideal Laboratory Types

  • Chemical laboratories

  • Pharmaceutical manufacturers

  • Petrochemical laboratories

  • Environmental testing laboratories

  • Academic chemistry departments

  • Government laboratories

  • Industrial quality assurance laboratories

Why Choose the VWR Explosion-Proof Laboratory Refrigerator?

The VWR Explosion-Proof Laboratory Refrigerator provides an important level of protection for laboratories that cannot safely use conventional refrigeration equipment. Explosion-proof electrical construction, hardwired installation, and a hermetically sealed refrigeration system help protect personnel and facilities when storing flammable chemicals in hazardous locations. Its 14 cubic foot capacity provides ample refrigerated storage while maintaining compliance with laboratory safety practices.

Typical Stored Materials

Typical stored materials include flammable solvents, alcohols, chromatography mobile phases, petroleum samples, volatile organic compounds, pharmaceutical intermediates, laboratory standards, chemical reagents, reference materials, and other hazardous chemicals requiring refrigerated storage.

VWR Explosion-Proof Laboratory Refrigerator Key Specifications

Identification

  • Manufacturer: VWR

  • Product Type: Explosion-Proof Laboratory Refrigerator

Capacity

  • 14 cu. ft. (396.4 L)

Temperature Range

  • 0°C to 10°C

  • Factory operating temperature approximately 4°C

Construction

  • Heavy-duty painted steel cabinet

  • Solid insulated door

  • One adjustable shelf and two fixed shelves

  • Non-sparking magnetic door gaskets

  • Keyed door lock

Refrigeration System

  • Hermetically sealed compressor

  • Explosion-proof electrical construction

  • Adjustable temperature controls

Safety Features

  • Explosion-proof design for hazardous locations

  • Hardwired installation

  • Suitable for refrigerated storage in hazardous environments

  • Designed in accordance with NFPA 45 and NFPA 70 recommendations

Electrical Requirements

  • 115V, 60 Hz

  • 4 Amps

  • Hardwired installation required

Exterior Dimensions

  • 32" W × 31" D × 66" H

Interior Dimensions

  • 27" W × 16" D × 51.5" H

Weight

  • 238 lbs. (108 kg)

Operating Costs & Energy Considerations

The VWR Explosion-Proof Laboratory Refrigerator is engineered primarily for hazardous-location safety rather than maximum energy efficiency. Routine condenser cleaning, proper ventilation, and periodic inspection of door gaskets will help maintain dependable refrigeration performance and extend equipment life.

Specifications compiled from VWR manufacturer literature and American Instrument Exchange product information.

Refurbishment & Testing Process

Every VWR Explosion-Proof Laboratory Refrigerator received by American Instrument Exchange undergoes a comprehensive inspection before shipment. Our technicians verify temperature accuracy using calibrated temperature monitoring equipment, inspect the refrigeration system, shelving, door gaskets, cabinet integrity, and overall operation. Each refrigerator is thoroughly cleaned, temperature tested, and verified for dependable laboratory performance before shipment.

Laboratories building complete hazardous-material storage systems frequently purchase chemical fume hoods, laboratory freezers, biosafety cabinets, and additional laboratory refrigerators from American Instrument Exchange.

Shipping & Installation Considerations

The VWR Explosion-Proof Laboratory Refrigerator ships by freight from American Instrument Exchange. Before installation, verify doorway clearances, available electrical service, and compliance with local electrical and fire codes. Because this is an explosion-proof refrigerator, it must be hardwired by a qualified electrician into an approved conduit box. Allow the refrigerator to stabilize at operating temperature before loading valuable laboratory materials.

Frequently Asked Questions

How much storage capacity does this refrigerator provide?

The refrigerator provides 14 cubic feet (396.4 liters) of refrigerated laboratory storage.

What temperature range does it maintain?

The refrigerator operates over a temperature range of 0°C to 10°C and typically cycles around 4°C for laboratory storage.

Does this refrigerator require hardwiring?

Yes. The unit is supplied without a power cord or plug and must be hardwired into an appropriate conduit box by a qualified electrician.

How is the refrigerator tested before shipment?

Every refurbished VWR Explosion-Proof Laboratory Refrigerator is inspected and temperature tested by American Instrument Exchange technicians. Refrigeration performance, temperature controls, shelving, door seals, cabinet integrity, and overall functionality are verified before shipment.

Testing Procedures for Used and Refurbished Laboratory Refrigerators
Ensure Temperature Stability and Cold Storage Accuracy

Used and refurbished laboratory refrigerators from American Instrument Exchange are thoroughly tested to ensure reliable cold storage performance, temperature accuracy, and compliance with laboratory safety and storage standards. Each unit undergoes a detailed inspection and functional verification to meet the requirements of clinical, pharmaceutical, research, and industrial labs.

Why Testing Used Laboratory Refrigerators Is Essential

Before a laboratory refrigerator is reintroduced into any laboratory setting, it must be carefully tested for:

• Temperature uniformity and stability
• Cooling system integrity and compressor function
• Door seal performance and insulation effectiveness
• Electrical safety and compliance with lab standards

Thorough testing minimizes the risk of temperature excursions, product loss, and system failures—especially critical for storing reagents, vaccines, biological samples, pharmaceuticals, and temperature-sensitive materials.

Our structured testing protocol ensures refurbished lab refrigerators deliver consistent, dependable performance across all makes and models.

Key Testing Procedures for Refurbished and Used Laboratory Refrigerators

1. Initial Visual Inspection
• Check exterior surfaces, hinges, and door gaskets for wear, corrosion, or damage
• Inspect power cords, plugs, and control panels for fraying or visible defects
• Verify that shelves, racks, and internal compartments are present and functional

2. Functional Testing
• Power on the refrigerator and verify that display screens, alarms, and controls function properly
• Test door latches and sealing mechanisms to ensure secure closure and insulation
• Run preliminary temperature cycles to observe startup behavior and cooling response

3. Temperature and Stability Testing
• Test internal temperature sensors and controllers using thermometers/probes
• Monitor temperature at multiple points within the chamber over an extended period
• Verify accuracy across user-defined setpoints (e.g., +4°C, +2°C to +8°C range)

4. Cooling System and Compressor Evaluation
• Test compressor startup, cycling, and operating noise levels
• Inspect condenser coils and fans for proper function and cleanliness
• Check refrigerant pressure (if applicable) and confirm there are no leaks or system inefficiencies

5. Electrical Safety and Compliance Checks
• Ensure all internal and external electrical components are properly grounded
• Perform insulation resistance and leakage current tests
6. Final Quality Assurance Checklist
• Run full operational tests over a 24-hour period
• Document temperature logs, control responsiveness, and system performance
• Complete a detailed service report

All refurbished laboratory refrigerators from American Instrument Exchange are backed by our rigorous testing standards and quality commitment. From vaccine storage to reagent preservation, you can count on consistent cold storage performance from day one.

Repair and Refurbishment Protocol for Used Laboratory Refrigerators
Cold Storage Performance from American Instrument Exchange

At American Instrument Exchange, every used laboratory refrigerator we offer is subjected to a comprehensive repair and refurbishment protocol. Whether it’s a general-purpose lab refrigerator, pharmaceutical-grade unit, or undercounter model, our goal is to restore each system to peak performance—ensuring consistent temperature control and reliable to laboratory storage standards.

Why Choose a Refurbished Laboratory Refrigerator from American Instrument Exchange?

• Fully tested for cold storage accuracy and operational safety
• Backed by a warranty
• Cost-effective alternative to new refrigeration equipment
• Sustainably refurbished to reduce environmental impact
• Serviced by expert technicians with decades of laboratory equipment experience

Step-by-Step Refurbishment Protocol

Every laboratory refrigerator from American Instrument Exchange follows a detailed, multi-stage process to ensure top-tier performance and reliability:

1. Initial Intake & Inspection

• Visual inspection for dents, corrosion, or structural damage
• Mechanical check of door hinges, gaskets, seals, and internal components
• Documentation of missing parts or areas requiring repair or replacement

2. Cleaning and Decontamination

• Internal and external cleaning using non-abrasive, lab-safe cleaners
• Decontamination/cleaning of all surfaces to remove biological or chemical residues
• Thorough cleaning of shelving, drawers, and removable interior parts

3. Mechanical and Electrical Repairs

• Replacement or repair of worn mechanical parts (door latches, seals, hinges, etc.)
• Inspection and restoration of control interfaces, thermostats, sensors, and displays
• Evaluation of wiring, relays, fuses, and power supply components for safety and integrity

4. Refrigeration System and Compressor Evaluation

• Testing of compressor function, condenser coils, and evaporators
• Inspection for refrigerant leaks and pressure performance (if applicable)
• Verification of proper airflow, fan operation, and defrost cycles

5. Temperature and Performance Testing

• Testing of temperature sensors and control boards
• Verification of temperature stability and uniformity throughout the chamber
• Testing of multiple setpoints (e.g., +2°C to +8°C) over extended operating periods

6. Safety and Compliance Checks

• Inspection of grounding and electrical insulation per lab safety standards
• Testing of alarm systems (if applicable) for temperature excursions or power failure

7. Final Quality Control and Documentation

• Multi-day operational test under real-world conditions
• Final cleaning, inspection, and performance validation
• Detailed service report with temperature logs, repair notes, and certification of refurbishment

Trust the Experts at American Instrument Exchange

With over 50 years of industry experience, American Instrument Exchange is a leading provider of high-quality refurbished laboratory equipment. Our proven laboratory refrigerator refurbishment protocol ensures consistent, accurate cold storage—so you can safeguard your samples, reagents, and research materials with confidence.

Warranty Information

All used equipment sold by American Instrument Exchange is refurbished/reconditioned, tested to meet specifications, cleaned, and comes with a 90-day warranty from the date of shipment unless otherwise specified. New products include a warranty from the manufacturer. Your warranty is only valid if you have paid your invoice. If your invoice is over 30 days old, warranty work will not be performed until your invoice is paid in full.  American Instrument Exchange is not liable for any special or consequential damages that result from the use of, or the inability to use, the materials on this site or the performance of the products,

Exceptions include international sales and items that are agreed to be sold with a lesser warranty or items that are agreed to be sold “as is.”

Should there be visible shipping damage to packaging or concealed shipping damage to item(s) supplied, buyer must communicate such damage to the delivering entity or carrier upon receipt first and to American Instrument the same day item is accepted / signed for. Failure to do so voids warranty. Customer decision to involve / procure third party or OEM service other than American Instrument or AIE assigned vendors to resolve in-warranty instrument failure or defect without first placing a service call to American Instrument voids warranty.

Return & Cancellation Policy

American Instrument Exchange will authorize the return of and issue a full refund for any equipment that is not working as represented and could not be replaced or repaired to the customer’s satisfaction by AIE or a representative of AIE during the warranty period. Returns without prior authorization and an RMA number will not be accepted. 

Customers are responsible for proper packaging of the equipment and all included accessories. American Instrument will pay for insurance and shipping costs. Non-inventory items bought specifically to fill customer orders may possibly be returned, but on a case by case basis.

Customers who cancel orders after refurbishing has begun or after delivery where there are no equipment performance issues may be subject to a Restocking/Cancellation Fee of 20% or more and responsible for shipping costs for both the delivery and return.

Ongoing and Extended Support Options

American Instrument Exchange is able to provide additional repair and support services outside the warranty terms for most products that sell.

For biosafety cabinets, clean benches, chemical fume hoods, and other clean air products our team at AIE Laboratory Services can provide certifications, repairs, decontaminations, and other services for customers throughout New England, and guidance and limited technical assistance for any customers in the US and Canada for new or refurbished biosafety cabinets purchased from American Instrument Exchange.

For all other products, the repair service team at American Instrument Exchange can provide both field and depot repair evaluations. We can easily schedule and coordinate the pick-up and transportation of your malfunctioning equipment. By offering hourly rates lower than most OEM service providers and eliminating travel time for technicians, we help reduce repair costs. Furthermore, our lab depot repair service minimizes or eliminates scheduling delays. Our experienced service technicians are qualified to restore your equipment to factory specifications.

Please contact our team at info@americaninstrument.com for more details.

FAQ

Does the equipment come with a warranty?

Yes, all equipment from American Instrument Exchange comes with a warranty.
Used/reconditioned/refurbished equipment is tested to make sure it operates to the manufacturer's specifications and comes with a 90-day warranty (unless otherwise noted). New equipment comes with a warranty from the manufacturer.

How do you test and recondition the equipment?

Testing requirements vary across all the equipment categories we sell. Temperature controlled units are tested with NIST traceable probes and data loggers. Gaskets, external housing, glass, handles, controls and control screens are all thoroughly inspected for integrity and functionality.
We test all equipment using the manufacturer's parameters and specifications to confirm that the unit operates properly.

How do products ship from American Instrument Exchange?

We can ship products both nationally and internationally using a well-established network of transporters. Smaller items will ship via UPS. Larger equipment will be palletized and shipped either LTL or air-ride, depending on the specific equipment and/or delivery requirements.

Return and order cancellation policy

American Instrument Exchange will accept the return of any product that is not working as represented and could not be replaced or repaired to the customer's satisfaction during the warranty period. Customers who cancel orders after refurbishing has begun will be subject to a cancellation and restocking fee of between 20% and 40%.

Are discounts available for educational or non-profit institutions, or for bulk purchases?

Yes – call or email us today with a list of equipment you're looking for and we can potentially put together a quote that includes a discount.

How does your price matching policy work?

We strive to keep our pricing competitive and reasonable. If you find comparable items for sale at a lower price from other vendors please send us a link and or quote for the item and we'll do our best to match or beat the price.

Can I see the equipment in person before committing to a purchase?

Customers are welcome to make an appointment to view equipment in our warehouse. Given our workload we may not be able to test or clean the equipment prior to your arrival if you have not committed to an order.

1023 Western Avenue is a working warehouse, not a showroom, and customers should dress appropriately and wear closed toe shoes.

Can I ship products that I purchased elsewhere to American Instrument for repair?

Yes – please see our repair services page for more information.