The Thermo Scientific Revco REC4504A21 Chromatography Refrigerator is a laboratory-grade chromatography refrigerator engineered for protein purification, FPLC systems, HPLC sample preparation, and other temperature-sensitive analytical workflows. Offering 45.8 cubic feet (1,297 L) of refrigerated storage and maintaining temperatures between 1°C and 8°C, the REC4504A21 provides the space and specialized infrastructure needed to operate complete chromatography systems while protecting valuable proteins, antibodies, enzymes, and biological samples.
Purpose-built for chromatography applications, the REC4504A21 incorporates dual 2-inch instrument access ports, an interior duplex electrical outlet, double sliding glass doors, positive forced-air circulation, and a microprocessor control system. These features allow chromatography pumps, fraction collectors, columns, and associated instrumentation to operate inside the refrigerated chamber while maintaining stable temperatures throughout lengthy purification runs.
Interior duplex electrical outlet with separate power cord
Positive forced-air circulation
Microprocessor temperature controller
Automatic demand defrost
Large digital display
Audible and visual alarm system
Battery backup
Remote alarm contacts
Adjustable epoxy-coated wire shelving
Heavy-duty industrial refrigeration system
Four casters (two locking)
Refurbished and performance tested by American Instrument Exchange
90-day warranty included.
Overview
The REC4504A21 was developed specifically for chromatography laboratories requiring refrigerated operation of complete purification systems. Rather than simply storing samples, this refrigerator accommodates chromatography pumps, fraction collectors, purification columns, and associated instrumentation while maintaining tightly controlled refrigerated conditions.
Two large access ports allow tubing, electrical cables, and monitoring probes to enter the cabinet without significantly affecting internal temperatures. The interior duplex outlet simplifies equipment installation, while the double sliding glass doors provide excellent visibility and conserve laboratory space during operation. Positive forced-air circulation maintains excellent temperature uniformity across all shelving levels, helping protect sensitive proteins during extended purification procedures.
Common Applications
The REC4504A21 is commonly used for ÄKTA chromatography systems, Bio-Rad NGC purification systems, FPLC, affinity chromatography, ion exchange chromatography, size exclusion chromatography, recombinant protein purification, monoclonal antibody purification, enzyme purification, fraction collection, refrigerated HPLC sample preparation, and storage of chromatography buffers, biological standards, media, and laboratory reagents.
Ideal Laboratory Types
Biotechnology laboratories
Pharmaceutical manufacturers
Protein purification laboratories
Structural biology laboratories
Bioprocess development laboratories
Molecular biology laboratories
University research laboratories
Government research laboratories
Why Choose the Thermo Scientific Revco REC4504A21?
The REC4504A21 combines the proven durability of the Revco platform with specialized chromatography features rarely found on conventional laboratory refrigerators. Its dual access ports, interior electrical outlet, positive airflow system, and spacious cabinet allow complete chromatography systems to operate under refrigerated conditions while maintaining excellent temperature stability and sample protection.
Typical Stored Materials
Typical contents include ÄKTA chromatography systems, Bio-Rad NGC purification systems, chromatography columns, fraction collectors, chromatography buffers, purified proteins, monoclonal antibodies, enzymes, biological standards, chromatography media, cell lysates, laboratory reagents, and other temperature-sensitive research materials.
Thermo Scientific Revco REC4504A21 Specifications
Specification
Details
Manufacturer
Thermo Scientific Revco
Model
REC4504A21
Product Type
High-Performance Chromatography Refrigerator
Temperature Range
1°C to 8°C
Storage Capacity
45.8 cu. ft. (1,297 L)
Exterior Dimensions (W × D × H)
56.5" × 36.6" × 79.1"
Interior Dimensions (W × D × H)
52.5" × 26" × 58"
Electrical Requirements
115V, 60Hz, 20 Amp
Door Style
Double sliding insulated glass doors
Shelving
Four full-width and two half-width adjustable epoxy-coated wire shelves
Specifications compiled from Thermo Scientific manufacturer literature and American Instrument Exchange product information.
Refurbishment & Testing Process
Every Thermo Scientific Revco REC4504A21 received by American Instrument Exchange undergoes a comprehensive inspection before shipment. Our technicians verify cabinet temperature using calibrated monitoring equipment, inspect the refrigeration system, controller, alarms, access ports, interior electrical outlet, shelving, door operation, and overall cabinet condition. Each refrigerator is thoroughly cleaned, temperature tested, and verified for dependable chromatography and laboratory operation before shipment.
The REC4504A21 ships by freight from American Instrument Exchange. Before installation, verify adequate electrical service, ventilation clearance, and sufficient floor loading capacity. Allow the cabinet to stabilize at operating temperature before installing chromatography equipment or loading valuable samples. Routing tubing and power cables through the factory-installed access ports helps maintain stable cabinet temperatures during operation.
Frequently Asked Questions
What makes the REC4504A21 different from a standard laboratory refrigerator?
The REC4504A21 is specifically engineered for chromatography applications and includes dual 2-inch access ports, an interior duplex electrical outlet, and a cabinet configuration that supports operating complete chromatography systems inside the refrigerated chamber.
Can an ÄKTA or Bio-Rad NGC system operate inside this refrigerator?
Yes. The REC4504A21 is designed to accommodate complete FPLC and chromatography systems, including pumps, columns, fraction collectors, and associated equipment operating under refrigerated conditions.
What temperature range does the refrigerator maintain?
The REC4504A21 is designed to maintain temperatures between 1°C and 8°C, with a factory preset of 4°C.
How is the refrigerator tested before shipment?
Every refurbished Thermo Scientific Revco REC4504A21 is inspected and temperature tested by American Instrument Exchange technicians. Refrigeration performance, temperature controls, alarms, access ports, interior electrical outlet, shelving, and overall cabinet functionality are verified before shipment.
CLICK HERE to see the Thermo Revco REC4504A glass door laboratory refrigerator on our YouTube channel
Testing Procedures for Used and Refurbished Laboratory Refrigerators Ensure Temperature Stability and Cold Storage Accuracy
Used and refurbished laboratory refrigerators from American Instrument Exchange are thoroughly tested to ensure reliable cold storage performance, temperature accuracy, and compliance with laboratory safety and storage standards. Each unit undergoes a detailed inspection and functional verification to meet the requirements of clinical, pharmaceutical, research, and industrial labs.
Why Testing Used Laboratory Refrigerators Is Essential
Before a laboratory refrigerator is reintroduced into any laboratory setting, it must be carefully tested for:
• Temperature uniformity and stability
• Cooling system integrity and compressor function
• Door seal performance and insulation effectiveness
• Electrical safety and compliance with lab standards
Thorough testing minimizes the risk of temperature excursions, product loss, and system failures—especially critical for storing reagents, vaccines, biological samples, pharmaceuticals, and temperature-sensitive materials.
Our structured testing protocol ensures refurbished lab refrigerators deliver consistent, dependable performance across all makes and models.
Key Testing Procedures for Refurbished and Used Laboratory Refrigerators
1. Initial Visual Inspection
• Check exterior surfaces, hinges, and door gaskets for wear, corrosion, or damage
• Inspect power cords, plugs, and control panels for fraying or visible defects
• Verify that shelves, racks, and internal compartments are present and functional
2. Functional Testing
• Power on the refrigerator and verify that display screens, alarms, and controls function properly
• Test door latches and sealing mechanisms to ensure secure closure and insulation
• Run preliminary temperature cycles to observe startup behavior and cooling response
3. Temperature and Stability Testing
• Test internal temperature sensors and controllers using thermometers/probes
• Monitor temperature at multiple points within the chamber over an extended period
• Verify accuracy across user-defined setpoints (e.g., +4°C, +2°C to +8°C range)
4. Cooling System and Compressor Evaluation
• Test compressor startup, cycling, and operating noise levels
• Inspect condenser coils and fans for proper function and cleanliness
• Check refrigerant pressure (if applicable) and confirm there are no leaks or system inefficiencies
5. Electrical Safety and Compliance Checks
• Ensure all internal and external electrical components are properly grounded
• Perform insulation resistance and leakage current tests 6. Final Quality Assurance Checklist
• Run full operational tests over a 24-hour period
• Document temperature logs, control responsiveness, and system performance
• Complete a detailed service report
All refurbished laboratory refrigerators from American Instrument Exchange are backed by our rigorous testing standards and quality commitment. From vaccine storage to reagent preservation, you can count on consistent cold storage performance from day one.
Repair and Refurbishment Protocol for Used Laboratory Refrigerators Cold Storage Performance from American Instrument Exchange
At American Instrument Exchange, every used laboratory refrigerator we offer is subjected to a comprehensive repair and refurbishment protocol. Whether it’s a general-purpose lab refrigerator, pharmaceutical-grade unit, or undercounter model, our goal is to restore each system to peak performance—ensuring consistent temperature control and reliable to laboratory storage standards.
Why Choose a Refurbished Laboratory Refrigerator from American Instrument Exchange?
• Fully tested for cold storage accuracy and operational safety
• Backed by a warranty
• Cost-effective alternative to new refrigeration equipment
• Sustainably refurbished to reduce environmental impact
• Serviced by expert technicians with decades of laboratory equipment experience
Step-by-Step Refurbishment Protocol
Every laboratory refrigerator from American Instrument Exchange follows a detailed, multi-stage process to ensure top-tier performance and reliability:
1. Initial Intake & Inspection
• Visual inspection for dents, corrosion, or structural damage
• Mechanical check of door hinges, gaskets, seals, and internal components
• Documentation of missing parts or areas requiring repair or replacement
2. Cleaning and Decontamination
• Internal and external cleaning using non-abrasive, lab-safe cleaners
• Decontamination/cleaning of all surfaces to remove biological or chemical residues
• Thorough cleaning of shelving, drawers, and removable interior parts
3. Mechanical and Electrical Repairs
• Replacement or repair of worn mechanical parts (door latches, seals, hinges, etc.)
• Inspection and restoration of control interfaces, thermostats, sensors, and displays
• Evaluation of wiring, relays, fuses, and power supply components for safety and integrity
4. Refrigeration System and Compressor Evaluation
• Testing of compressor function, condenser coils, and evaporators
• Inspection for refrigerant leaks and pressure performance (if applicable)
• Verification of proper airflow, fan operation, and defrost cycles
5. Temperature and Performance Testing
• Testing of temperature sensors and control boards
• Verification of temperature stability and uniformity throughout the chamber
• Testing of multiple setpoints (e.g., +2°C to +8°C) over extended operating periods
6. Safety and Compliance Checks
• Inspection of grounding and electrical insulation per lab safety standards
• Testing of alarm systems (if applicable) for temperature excursions or power failure
7. Final Quality Control and Documentation
• Multi-day operational test under real-world conditions
• Final cleaning, inspection, and performance validation
• Detailed service report with temperature logs, repair notes, and certification of refurbishment
Trust the Experts at American Instrument Exchange
With over 50 years of industry experience, American Instrument Exchange is a leading provider of high-quality refurbished laboratory equipment. Our proven laboratory refrigerator refurbishment protocol ensures consistent, accurate cold storage—so you can safeguard your samples, reagents, and research materials with confidence.
Warranty Information
All used equipment sold by American Instrument Exchange is refurbished/reconditioned, tested to meet specifications, cleaned, and comes with a 90-day warranty from the date of shipment unless otherwise specified. New products include a warranty from the manufacturer. Your warranty is only valid if you have paid your invoice. If your invoice is over 30 days old, warranty work will not be performed until your invoice is paid in full. American Instrument Exchange is not liable for any special or consequential damages that result from the use of, or the inability to use, the materials on this site or the performance of the products,
Exceptions include international sales and items that are agreed to be sold with a lesser warranty or items that are agreed to be sold “as is.”
Should there be visible shipping damage to packaging or concealed shipping damage to item(s) supplied, buyer must communicate such damage to the delivering entity or carrier upon receipt first and to American Instrument the same day item is accepted / signed for. Failure to do so voids warranty. Customer decision to involve / procure third party or OEM service other than American Instrument or AIE assigned vendors to resolve in-warranty instrument failure or defect without first placing a service call to American Instrument voids warranty.
Return & Cancellation Policy
American Instrument Exchange will authorize the return of and issue a full refund for any equipment that is not working as represented and could not be replaced or repaired to the customer’s satisfaction by AIE or a representative of AIE during the warranty period. Returns without prior authorization and an RMA number will not be accepted.
Customers are responsible for proper packaging of the equipment and all included accessories. American Instrument will pay for insurance and shipping costs. Non-inventory items bought specifically to fill customer orders may possibly be returned, but on a case by case basis.
Customers who cancel orders after refurbishing has begun or after delivery where there are no equipment performance issues may be subject to a Restocking/Cancellation Fee of 20% or more and responsible for shipping costs for both the delivery and return.
Ongoing and Extended Support Options
American Instrument Exchange is able to provide additional repair and support services outside the warranty terms for most products that sell.
For biosafety cabinets, clean benches, chemical fume hoods, and other clean air products our team at AIE Laboratory Services can provide certifications, repairs, decontaminations, and other services for customers throughout New England, and guidance and limited technical assistance for any customers in the US and Canada for new or refurbished biosafety cabinets purchased from American Instrument Exchange.
For all other products, the repair service team at American Instrument Exchange can provide both field and depot repair evaluations. We can easily schedule and coordinate the pick-up and transportation of your malfunctioning equipment. By offering hourly rates lower than most OEM service providers and eliminating travel time for technicians, we help reduce repair costs. Furthermore, our lab depot repair service minimizes or eliminates scheduling delays. Our experienced service technicians are qualified to restore your equipment to factory specifications.
Yes, all equipment from American Instrument Exchange comes with a warranty. Used/reconditioned/refurbished equipment is tested to make sure it operates to the manufacturer's specifications and comes with a 90-day warranty (unless otherwise noted). New equipment comes with a warranty from the manufacturer.
How do you test and recondition the equipment?
Testing requirements vary across all the equipment categories we sell. Temperature controlled units are tested with NIST traceable probes and data loggers. Gaskets, external housing, glass, handles, controls and control screens are all thoroughly inspected for integrity and functionality. We test all equipment using the manufacturer's parameters and specifications to confirm that the unit operates properly.
How do products ship from American Instrument Exchange?
We can ship products both nationally and internationally using a well-established network of transporters. Smaller items will ship via UPS. Larger equipment will be palletized and shipped either LTL or air-ride, depending on the specific equipment and/or delivery requirements.
Return and order cancellation policy
American Instrument Exchange will accept the return of any product that is not working as represented and could not be replaced or repaired to the customer's satisfaction during the warranty period. Customers who cancel orders after refurbishing has begun will be subject to a cancellation and restocking fee of between 20% and 40%.
Are discounts available for educational or non-profit institutions, or for bulk purchases?
Yes – call or email us today with a list of equipment you're looking for and we can potentially put together a quote that includes a discount.
How does your price matching policy work?
We strive to keep our pricing competitive and reasonable. If you find comparable items for sale at a lower price from other vendors please send us a link and or quote for the item and we'll do our best to match or beat the price.
Can I see the equipment in person before committing to a purchase?
Customers are welcome to make an appointment to view equipment in our warehouse. Given our workload we may not be able to test or clean the equipment prior to your arrival if you have not committed to an order.
1023 Western Avenue is a working warehouse, not a showroom, and customers should dress appropriately and wear closed toe shoes.
Can I ship products that I purchased elsewhere to American Instrument for repair?
Yes – please see our repair services page for more information.