Thermo Sorvall Legend 21R Refrigerated Microcentrifuge
The Thermo Sorvall Legend 21R Refrigerated Microcentrifuge can accelerate from a stopped position to 21,0000 x g in just 12 seconds and cooling from ambient to -9°C in under ten minutes.
Thermo Sorvall Legend 21R features
- Max speed: 14,800rpm
- Max RCF: 21,100 x g
- Includes rotor; call or email for details
- 90-day warranty
The Thermo Sorvall Legend 21R is compatible with a variety of rotors (sold separately – call for details) and accommodates 36 x 0.5mL microtubes, 24 x 2 mL tubes, or 8 x 8 PCR strips.
Rotors feature ClickSeal biocontainment lids that click into place, ensuring securely contained samples. The Thermo Legend 21R is also compact and the space-saving design has a small footprint so that the microcentrifuge doesn’t take up valuable bench space in the lab.
Thermo Sorvall Legend 21RTesting
Most of our equipment is used and reconditioned. Our technicians test the performance of all centrifuges. We check for proper balance on the rotating assembly. The bearings, seals and gaskets on the spindle assembly are in good shape, and the controls work properly.
If the centrifuge is refrigerated we also test to make sure that the machine reaches and hold temperature properly. We measure all tolerances. We clean the exterior and interior of all units. If the unit is being purchased with a rotor we test the rotor/centrifuge combination together.
Thermo Sorvall Legend 21R Specifications
|
Speed Range
|
to 14,800 rpm
|
|
Capacity
|
24 x 1.5/2.0mL
|
|
Temperature Range
|
-9°C to +40°C
|
Pictures
The photos above represent our equipment currently stored in our warehouse. Since we often stock multiple units of each model, we may not feature photos of every unit.However, if you request, we will provide photos of your specific equipment after we test and clean it.
For inquiries about the Thermo Sorvall Legend 21R, call us at 978-521-2221 or fill out the form on this page.
Thermo Scientific Micro 21R Centrifuge – Frequently Asked Questions
What applications is the Thermo Micro 21R used for?
Common applications include DNA/RNA extraction, protein lysate preparation, PCR and qPCR workflows, and clinical and research sample prep.
Installation & Setup
Does the Micro 21R require special installation?
No, installition requirements are minimal and easy to meet. Basically this is a plug-and-play benchtop unit. The only requirements for use are:
- Placement on a stable, level surface
- Adequate ventilation space
- Connection to the correct electrical supply
How does the refrigeration system work?
The Micro 21R includes an integrated refrigeration unit that rapidly cools samples and maintains set temperature during runs. It uses standard and easy to find/replace refrigerant such as R-134a.